ISO/TC 194/WG 12 N15

February 1999

1st Working Draft Revision of ISO 10993-12:1996, Biological evaluation of medical devices - Part 12: Sample Preparation and Reference Materials (6 February 1999)



Introduction

This International Standard specifies methods of sample preparation and use of reference materials in biological evaluation. Because of the many different biological assay systems described in ISO 10993, the individual standards should be consulted to ascertain the appropriateness of these recommendations for a specific test system.

Sample preparation methods should consider both the biological evaluation methods and the materials being evaluated. Each biological test restricts selection of solid samples and extraction solvents or conditions by its own methodology.

This part of ISO 10993 is based on existing national and international specifications, regulations and standards wherever possible. It is open to regular review whenever new research work is presented to improve the state of scientific knowledge.

1 Scope

This part of ISO 10993 specifies requirements and gives guidance on procedures to be followed in the preparation of samples of medical devices for testing in biological systems in accordance with one or more parts of the ISO 10993 series. These include;

2 Normative references Normative references

The following standards contain provisions that, through reference in this text, constitute provisions of this international Standard. At the time of publication, the editions indicated were valid. All standards are subject to revision, and parties to agreements based on this International Standard are encouraged to investigate the possibility of applying the most recent editions of the standards indicated below. Members of international Electrotechnical Commission (EC) and international Organization for Standards (ISO) maintain registers of currently valid international Standards.

ISO/IEC Directives, Part 2: 1992, Methodology for the development of international Standards,
Annex B - Mention of reference materials

3 Definitions

3.1 blank: Extraction vehicle not containing the specimen used for comparison with the extract liquid.

3.2 certified reference material (CRM): Reference material, accompanied by a certificate, one or more of whose property values are certified by a procedure which establishes its traceability to an accurate realization of the unit in which the property values are expressed, and for which each certified value is accompanied by an uncertainty at a stated level of confidence (see ISO Guide 30).

NOTE: standard reference material (SRM): A trademark protected certification supplied by the National Institutes for Standards and Technology, Gaithersburg. MD, USA.

3.3 extract liquid: Liquid that is tested for biological response after it has been used for extraction of the device.

3.4 homogeneous: The condition of being of uniform structure or composition with respect to the biological endpoint under study. The RM is said to be homogeneous if the biological response to a specific test is found to lie within the specified uncertainty limits of the test, irrespective of the site in the batch or lot of material from where the test sample is taken.

3.5 negative control: A negative control is any well characterized material or substance, which when tested by the procedure described, demonstrates the suitability of the procedure to yield a reproducible, appropriate negative, nonreactive or background response in the test system.

3.6 positive control: A positive control is any well-characterized material or substance, which when tested by the procedure described, demonstrates the suitability of the procedure to yield a reproducible, appropriate positive or reactive response in the test system.

3.7 reference material (RM): A material or substance, one or more of whose property values are sufficiently homogeneous and well established to be used for the calibration of an apparatus, the assessment of a measurement method, or for assigning values to materials (see ISO Guide 30). For the purposes of the standard, a reference material is any well-characterized material or
substance, which when tested by the procedure described, demonstrates the suitability of the procedure to yield a reproducible, predictable response. The response may be negative or positive.

3.8 reference method: A thoroughly investigated test method, clearly and exactly describing the necessary conditions and procedures, for the evaluation of a specific biological endpoint, that has been shown to have accuracy and precision commensurate with its intended use and that can, therefore, be used to characterize a RM (see ISO Guide 30).

3.9 Stability of property values: The ability of a material, when stored under specified conditions, to maintain the stated biological response, within specified limits, for a specific period of time (see ISO Guide 30).

3.10 test material: Material, device, device portion, or component thereof subject to biological or chemical testing.

3.11 test sample: Test material or extract subject to biological or chemical testing.

4 Experimental controls

Experimental controls shall be used in biological evaluations to validate a test procedure and/or to evaluate the results from a new material. Depending on the biological test being used, negative and/or positive controls and blanks shall be used as appropriate. The same control may be applicable to different tests to allow cross-reference to established materials and test methods. Additional guidance on the selection of experimental controls is given in Annex B.

5 Use of reference materials as experimental controls

5.1 Reference or certified reference materials shall be used in biological tests as control materials to demonstrate the suitability of the procedure to yield a reproducible, predictable response, e.g. positive or negative. Use of a reference material in this way will ensure the comparability of the response between laboratories. The property values of any material used in this way shall be characterized with each biological test procedure for which the use of the material is desired. A material characterized and then certified for one reference test method or response, e.g. sensitization, shall not be used as a reference material for another, e.g. cytotoxicity, without additional characterization.

5.2 Reference materials used as experimental controls shall meet the established quality assurance procedures of the manufacturer and test laboratory. They shall be identified as to source, manufacturer, grade, and type. Reference materials used as experimental controls shall be fabricated, finished and their surfaces cleaned and sterilized in a manner appropriate to the projected application in human subjects. Where possible, the above processes should be identical to the processes employed in the manufacture of the finished device.

5.2.1 Reference materials used as experimental controls shall be in the same physical class as the test sample, i.e., polymer, ceramic, metal, colloid, etc. Pure chemicals may be used as Reference Materials for mechanistically based test procedures, e.g. mutagenicity and immune sensitization.

5.3 A comparable, clinically accepted device will satisfy the requirement of 5.2.

6 Characteristics of reference materials

6.1 Material screening versus medical device biocompatibility

The uses of the reference materials described in this document are limited to biological screening testing of the materials intended for use in the manufacture of medical devices. However, while not intended for the purpose, they are often used in the performance assessment of the finished device. The vertical standard for the device, when available, must address the biological testing of the product in the performance environment of the device. Biological testing described in the vertical standard takes precedent over testing performed to screen the materials for suitability.

6.2 Misuse of CRMs

The attention of the users of this standard is directed to the discussion of "proper use" and "misuse" of CRMs in the introduction to ISO Guide 33. This discussion points out areas of both potential under and over utilization of RMs and CRMs. Users of this standard should also note that the use of calibration materials to evaluate the biological response of materials under investigation within a single laboratory is acceptable.

6.3 One or more property values

RMs or CRMs used to determine the biological response in this way must be evaluated with each biological test procedure for which the use of the material is desired. It is not sufficient to qualify a material for one type of reference test method or response, e.g. sensitization, and declare it a reference material for another, e.g. cytotoxicity, without additional qualification testing.

6.4 Long term availability of the material

During the development of this standard, concern was expressed about the future availability of reference materials for determination of biological response. It is desirable for users to obtain a commitment of not less than 5 years from suppliers of the RMs or CRMs. A second, but less desirable, option is the publication by the source of the RM or CRM of an "open formulation" for the material, i.e. publication of the source materials and details of the processing needed to insure uniform batches of RM.

6.5 Certification of reference materials for biological safety testing

6.5.1 Individual laboratories certify reference materials. The individual laboratory determines the extent of chemical, physical and biological characterization. Commercially available articles may be used as reference materials (see Clause 5.3).

6.5.2 Certification of a reference material is a procedure that establishes the value of the biological response of the material under the test conditions specified, ensuring a traceability of the response, which leads to the issuance of a certificate. The biological response of the material should be established through interlaboratory tests.

7 Test material selection

7.1 It is preferable to test medical devices in their final product form whenever practical. When that is not possible, the second choice is representative portions of the device (see Clause 9). When these are not possible or practical, representative molded or extruded test specimens of the formulated material that have been preconditioned by the same processing as the final product shall be tested.

7.2 The same test material selection procedure applies when an extract is required.

8 Test sample and reference material preparation

8.1 Test samples and reference materials shall be handled to prevent contamination. Any residues from the manufacturing, fabrication, cleaning, sterilization, etc., processes, if present, should be considered to be integral to the device, device portion or component. Additional guidance on the preparation is given in Annex C.

8.1.1 Test samples from sterilized devices and reference materials shall be handled aseptically if appropriate to the test procedure.

8.1.2 Test samples from devices which are normally supplied non-sterile but are sterilized before use shall be sterilized by the method recommended by the manufacturer and handled aseptically if appropriate to the test procedure.

8.1.3 If test samples are cleaned prior to sterilization, the influence of the cleaning process and cleaning agent shall be considered.

8.2 If sterile test samples are required for the test procedure, the effect of the sterilization or resterilization process on the test sample and reference materials shall be considered.

8.3 When test samples and reference materials need to be cut into pieces, the influence of previously unexposed surfaces, e.g. lumens or cut surfaces, shall be considered. Techniques used for cutting medical devices into representative portions for testing should be as clean as practical to prevent contamination.

9 Selection of representative portions from a device Selection of representative portions from a device

9.1 If a device cannot be tested as a whole, each individual material in the final product shall be represented proportionally in the test sample.

9.1.1 The test sample of devices with surface coatings shall include both coating material and the substrate.

9.1.2 The test sample shall include a representative portion of the joint and/or seal if adhesives, radio frequency (RF) seals, or solvent seals are used.

9.1.3 Composite materials shall be tested as finished materials.

9.1.4 Where extracts are used in the test methods for evaluation of materials that cure in-situ, e.g. cements, adhesives and monomers, initiation of the extraction shall occur after the specified minimum cure that may occur during clinical use. For test methods that use these materials directly, e.g. direct contact or agar overlay cytotoxicity, implantation, some genotoxicity tests, and direct contact hemolysis, the material shall be used as in clinical use, with in-situ cure in the test system. Modification of the delivery system so that the designated size or weight of the material is delivered is appropriate.

9.2 When different materials are present in a single device, the potential for synergies and interactions shall be considered in the choice of test sample.

9.3 The test sample may be chosen to maximize the exposure of the test system to the components where components of a device are known to have a potential for a biological response.

10 Preparation of extracts of samples

If extracts of the device are required for a test protocol, the extraction media and conditions of extraction used shall be appropriate to the nature and use of the final product. Additional guidance on the extraction of samples is given in Annex D.

10.1 Containers for extraction.

10.1.1 The extraction shall be performed in clean, chemically inert closed containers with minimum headspace.

10.1.2 To ensure that the extraction vessels do not adulterate the extract of the test materials the extraction vessels shall be:

10.1.3 Extractions shall be performed under conditions which prevent contamination of the sample.

10.2 Extraction conditions.
Extraction is a complex process influenced by time, temperature, surface-area-to-volume ratio, extraction medium and the phase equilibrium' of the material. The effects of higher temperatures or other conditions on extraction kinetics and the identity of the extractant(s) shall be considered carefully if accelerated or exaggerated extraction is used. Based on current practices, standard conditions that have been used to provide a measure of the hazard potential of the device or material are described. Other conditions that simulate the extraction that occurs during clinical use or provide an adequate measure of the hazard potential may be used and shall be described and justification provided.

10.2.1 Standard extraction temperatures and times (with permissible ranges) are as follows (see also Clause D3): 10.2.2 Standard surface area, i.e., projected area, excluding indeterminate surface irregularity, and extractant volumes are:


Thickness

(mm)

Extraction ratio

(surface area/volume)

+-10%

Example of

Materials

< 0.5 6cm2/ml film, sheet,

tubing wall

0.5 - 1.0 3cm2/ml tubing wall, slab,

molded items

> 1.0 1.25cm2/ml natural elastomer
Not applicable 0.2g sample/ml

6cm2/ml

powder, pellets

foam, non-absorbent

---------------------------
1 The phase equilibrium of a material at a temperature controls the relative amounts of amorphous and crystalline phases present. For the amorphous phase, the glass transition temperature, Tg, dictates the polymer chain mobility and the diffusion rate in the phase. Usually, the diffusion rate is considerably higher above the Tg compared with that below. The diffusion rate is lowest in the crystalline phase. The extraction conditions shall not alter the phase equilibrium of the material. Phase alteration may affect the amount and type of extractables.
---------------------------


NOTE: While there are no standardized methods available at present for testing absorbents and hydrocolloids, the following is a suggested protocol: Determine the "absorption capacity" of the material, i.e. the amount of extractant absorbed by one gram of the material. The test sample should be 2 grams of the material. The extract volume should be 20 ml more than the "absorption capacity" of the 2 gram sample.

10.2.2.1 Other surface area to volume extraction ratios, e.g. those related to evaluation of porous surfaces, may be used provided that they simulate the conditions during clinical use or result in a measure of the hazard potential.

10.2.2.2 Elastomers, coated materials, composites, laminates, etc., should be tested intact whenever possible because of differences in extraction characteristics from the intact and cut surfaces.

10.2.2.3 Materials should be cut into small pieces before extraction to enhance submersion in the extract media, except when otherwise inappropriate. See, for example, Clause 10.2.2.2. For polymers, pieces approximately 10 mm x 50 mm or 5 mm x 25 mm are appropriate.

10.2.3 Examples of extraction media are 10.2.4 Extractions are commonly performed under static conditions. When agitation is considered to be appropriate, the method should be specified and reported.

10.2.5 Liquid extracts shall, if possible, be used immediately after preparation to prevent sorption onto the extraction container or other changes in composition. If an extract is stored longer than 24 hours, then the stability of the extract under the conditions of storage should be verified.

10.2.6 The extract should not routinely be processed by filtration, centrifugation or other methods to remove suspended particulates. However, if such processing is necessary, the rationale must be presented.

11 Records

Documentation of sample preparation shall include, but not be limited to:

  1. Source of material, device, device portion or component;
  2. Lot or batch number, where appropriate;
  3. Description of processing, cleaning or sterilization treatments, if appropriate;
  4. Extraction techniques, as appropriate, including documentation of the conditions for extraction.

Annex A - Bibliography (Informative)


ANNEX B - Experimental controls (Informative)

B.1 The materials listed in the following paragraphs may meet the criteria for an appropriate experimental control in selected tests. It is the responsibility of the investigator to make the appropriates choices. See Table B 1.

Table B1 - Summary of reference materials and controls

TEST

POSITIVE

CONTROL*

NEGATIVE

CONTROL*

REFERENCE

MATERIAL*

IMPLANTATION

PVC-org. Sn

PE

 

 

 

 

SPU-ZDEC

Silicone

 

 

 

 

Latex

Alumina

 

 

 

 

 

 

Stainless steel

 

 

CYTOTOXICITY

PVC-org. Sn

PE

 

 

 

 

SPU-ZDEC

 

 

 

 

 

 

SPU-ZBEC

 

 

 

 

 

 

Latex

 

 

 

 

BLOOD

COMPATIBILITY

 

 

 

 

PVC 7506

PUR 2541

 

 

 

 

 

 

 

 

* Abbreviations on this Table refer to specific materials available from sources designated in Clauses B.1 and B.2.

B.2 Examples of solid reference materials which have been used as negative controls are, for example, high density polyethylene(2, low density polyethylene(3,4, silica-free polydimethylsiloxane(4,5, polyvinylchloride(6, polyetherurethane(7, polypropylene(8, medical grade latex, aluminum oxide ceramic rods, stainless steel and titanium alloys. This information is given for the convenience of the user of this part of ISO 10993 and does not constitute an endorsement by ISO of the product.

B.3 Examples of materials which have been used as positive controls for solid samples are polyvinylchloride containing organo-tin additives(9, segmented polyurethane films containing zinc diethyl- or dibutyl-dithiocarbamate, certain latex formulations and solutions of zinc salts(10. Substances which have been used as positive controls for extract samples are dilutions of phenol and water. This information is given for the convenience of the user of this part of ISO 10993 and does not constitute an endorsement by ISO of the product.

-----------------------
2 High density polyethylene (Negative Control Plastic RS) can be obtained from the US Pharmacopeia (Rockville, MD USA).
3 PE 140 tubing is available from Rehau AG, D-8673 Rehau, Germany. PE film is available from Hoechst AG, D6230 Frankfurt 80, Germany.
4 Biomaterials Program, Devices and Technology Branch, National Heart, Lung and Blood Institute, NIH, 312 Federal Building, 7550 Wisconsin Ave., Bethesda, MD 20892, USA.
5 SIK 8363 tubing is available from Rehau AG, D-8673 Rehau, Germany.
6 PVC 7506 and PVC 7536 tubing is available from Rehau AG, D-8673 Rehau, Germany. PVC-DEHP and PVCTEHTM film is available from Hoechst AG, D-6230 Frankfurt 80, Germany.
7 PUR 2541 tubing is available from Rehau AG, D-8673 Rehau, Germany. PU film is available from Frontline Filmblasning, S-60003 Norrkoping, Sweden.
8 PP 146 tubing is available from Rehau AG, D-8673 Rehau, Germany. PP film is available from Hoechst AG, D6230 Frankfurt 80, Germany.
9 Positive control RS can be obtained from the US Pharmacopeia, Rockville, MD, 20852, USA.
10 AS 2696-1989 Medical Equipment - Single use urethral catheters (sterile) for general medical use.
-----------------------


ANNEX C - Principles of test material preparation (Informative)

This section provides general principles on and practices of material preparation and sample selection.

C.1 The material used in the biological assay should be representative of the composition, processes and surface characteristics of the final product. See Clause 7.

C.1.1 In the case of plastic and rubber materials, the composition should include resin, polymer and any additives. Alternate components of a formulation should be evaluated for substitution. The formulation description should specify the history of the material, e.g., thermal, virgin or regrind and the maximum allowable regrind.

C.1.2 Materials that may be re-sterilized by the same or alternative methods should be tested after treatment by the multiple sterilizations. For example, a material that is sterilized by radiation and re-sterilized by ethylene oxide should be tested after (a) irradiation and (b) irradiation plus ethylene oxide. If a "worst case" exposure can be identified with appropriate justification, testing may be performed after exposure to this treatment.

C.1.3 Ideally, all biological tests on a material should have the surface of the material exposed (as opposed to the bulk phase) to the cellular/biological environment. An alternative method to cutting the surface is fabrication of miniatures of the device using the same process (extrusion, dipping, etc.), temperatures, time, atmosphere, release agents, annealing, curing, cleaning, sterilization, etc., processes used in the manufacture of the device.

C.1.4 Metals used in biological tests should be from the same stock material used to fabricate the device and using the same machining, grinding, polishing, cleaning, passivation, surface treatment and sterilization used in the manufacture of the final product.

C.1.5 Ceramic materials used in biological tests should be manufactured from the same powder stock using the same casting, investing, molding, sintering, surface finishing and sterilization processes used to manufacture the device.

C.1.6 Bioprosthetic materials should be tested after they have been preserved under the manufacturers maximum and minimum allowable fixation times to allow for varying penetration of the fixative.

C.2 Extraction conditions that may cause particle generation should be considered in the design of tests on the material.

C.3 The amount of material and surface area thereof shall be appropriate to the biological and physical constraints of the test system. In practice, the use of a standard sample size for a specific assay is recommended.


ANNEX D - Principles for extraction of samples (Informative)

The purpose of an extract of a medical device is to provide a suitable test sample for determining the biological reactivity of any leachable substances in a biological system and to demonstrate the hazard potential for the use of the device in humans. If extracts of the device are prepared, the medium and conditions of extraction used should be appropriate to the nature and use of the final product as well as to the predictability (test rationale, sensitivity, etc) of the test method. Extraction conditions, therefore, should ideally reflect not only actual in-use conditions of the products but also predictability of the tests.

D.1 This standard assumes that the amount of extractable substance(s) is/are related to the period of extraction, the temperature, the ratio of surface-area-of-material to volume-of-extractant and the nature of the solvent.

D.2 The period of extraction should be sufficient to maximize the amount of material extracted. In practice, standard time and temperature conditions are recommended in lieu of specific chemical analyses. An alternative practice is repeated extraction followed by concentration to obtain sufficient extractable substance(s).

D.3 The extraction temperature should maximize the amount of extractable substances as well as simulate any extremely high temperatures the device may incur during clinical use. This simulation should not initiate significant degradation of the material. The extraction temperature is dependent upon the physicochemical characteristics of the device material(s). For example, for polymers, the extraction temperature chosen should be below the glass transition temperature. If the glass transition temperature is below the use temperature, the extraction temperature shall be below the melting temperature. Recommended conditions are shown in Section 10.2.1.

The following examples are presented to illustrate the interpretation of this section:

D.4 The ratio of the surface area of the device to the volume of extractant or solvent should be sufficient to:

In practice, the use of a standard area and solvent volume is recommended as described in Section 10.2.2 in lieu of device specific parameters. Some test methods require concentration of extracts to increase the sensitivity of the test.

NOTE: Concentration of extracts may result in loss of volatile materials such as ethylene oxide.

D.5 The solvent(s) selected as extractants shall:

In practice, the use of standard polar and non-polar solvents are recommended in Section 10.2.3 in lieu of device specific solvents.

NOTE: Standardization of the parameters given in Clauses D4 and D5 permit the use of data obtained from biological tests of medical devices for other types of applications, e.g. for the estimation of risk and to develop standardized databases.